PSL-100 Electron Lifetime & Diffusion Measurement System

Traditionally Dye-Sensitized Solar Cell (DSC) materials could only be tested by measuring the electrical performance of assembled solar cells. However in order to improve the conversion rate of the solar cells, it is necessary to improve short circuit current (Isc) and open circuit voltage (Voc). Therefore for DSC, it is very important to understand the characteristics of the electron transfer inside the cell, which significantly contribute to the performance of the cell.

With the PSL-100 Electron life time tester, the electron transfer process inside the DSC porous electrode can be determined and will give a more detailed guideline for material development and cell efficiency improvement.

The Electron Lifetime & Diffusion Measurement System “PSL-100” measures electron transfer characteristics, “Electron Diffusion Coefficient (D)” and “Electron Lifetime (τ)” for sample size level DSC based on the measurement theory of SLIM-PCV(Stepped Light-induced Transient Measurements of Photocurrent and Voltage method.

Electron Diffusion Length

"Electron Diffusion Length" is defined as the distance of transferring electron which is excited by light irradiance till it gets recombined. In other words, if this distance is shorter than the thickness of the DSC porous semiconductor electrode, all by light excited electrons cannot be extracted. So the Electron Diffusion Length(L) can be expressed as following by using Electron Diffusion Coefficient (D)and Electron Lifetime (τ): L=√D.τ

Features

  • Electron Lifetime.Electron Diffusion.Electron Density.I-V Curve measurement
  • Advanced system for DSC evaluation
  • For material development of DSC cells
  • Laser operated 660nm / 473nm
ASK A QUICK QUESTION
PSL-100 Electron Lifetime & Diffusion Measurement System

Specifications

Specifications  
Installation Condition Indoor use only
Operating Temperature +10°c ~ 40°c
Humidity Less than 70% (no condensation)

Measurement Unit PSL-100-M

General Specification  
Power Supply AC 100[V] ±10%
Frequency Range 50[Hz] / 60[Hz]
Power Consumption Less than 25[VA]
Warm-up Time More than 30 minutes
Outer Dimensions 480(W)x190(H)x350(D)[mm]
Without Handle and Rubber Legs
Weight 10 kg
Measurement Part Specification  
Electron Diffusion Coefficients Measurement  
Measurement Method Current measurement (Current/Voltage Conversation)
Conversion Ratio 1.0000[mA] / 1.0000[V]
Measurable Rabge -0.3000[mA] ~ +1.2000[mA]
A/D Conversion Rate 1:1
Value Data Rate ±0.005[mA]
Electron Lifetime Measurement  
Measurement Method Voltage measurement
Input Impedance More than 109[Ω]
Measurable Range -0.3000[V] ~ +1.2000[V]
A/D Conversion Rate 1:1
Value Data Error Within ±0.005[V]
Electron Density Measurement  
I-V Curve Measurement  
Control Method Bias Potential Applied Method
Control Output Impedance Less than 20[Ω]
Control Output Setup Range -0.3000[V] ~ +1.2000[V]
Control Output Setup Resolution 0.0001[V]
Control Output Setup Error ±0.005[V]
Voltage Measurement Unit  
Measurement Method Voltage measurement
Input Impedance More than 109[Ω]
Voltage measurement conversion ratio 1:1
Voltage Measurable Range -0.3000[V] ~ +1.2000[V]
A/D Conversion Rate 1:1
Value Data Error Within ±0.005[V]
Current Measurement Unit  
Measurement Method Current measurement (Current/Voltage Conversion)
Measurable Range 4 ranges(100mA, 10mA, 1mA, 100μA)
Conversion Ratio Current Ranges F.S. / 1.0000[V]
Measurable Ranges 100mA Range :-30.000[mA]~+100.00[mA]
10mA Range:-3.0000[mA]~+10.000[mA]
1mA Range:-0.3000[mA]~+1.0000[mA]
100μA Range:-30.000[μA]~+100.00[μA]
A/D Conversion Rate 1:1
Value Data Error 100mA Range:Within ±1.0[mA]
10mA Range :Within ±0.05[mA]
1mA Range : Within ±0.005[mA]
100μA Range: Within ±0.5[μA]
Communication Specification  
Communication Method Serial Communication Method
Communication Port USB port or RS-232C Port
Recommend to use USB Port
Communication Speed 19200bps

Optical Bench PSL-100-O

General Specification  
Outer Dimensions 803(W)×290(H)×303(D)[mm]
Without Rubber Legs
Weight 50[kg]
Optical Parts (Main Parts)  
Laser Specification (Standard Light Source)  
Laser Used MLX-D12-640-40 By Kiko
Safety Class 3B
Optical Lens φ6.4 Aspheric Glass Lens
Laser Light Wavelength 660[nm] ±10[nm]
Plano-Convex Lens Specification  
Lens SLB-30-60P By Sigma Koki
Outer Dimension φ30[mm] +0~-0.1[mm]
Focal Point Distance 60[mm] ±1%

Laser Operation Unit PSL-100-L

General Specification  
Outer Dimensions 150(W)×180(H)×280(D)[mm]
Without Rubber Legs
Weight 3.0[kg]
Power Supply AC100V ±10% or AC200V ±10%
Frequency Range 50Hz / 60Hz
Power Consumption Less than 6.3VA
Warm-up Time More than 30 minutes

Software PSL-100

Operating Environment  
OS MS Windows 7
HHD Capacity Approx. 10GB
Memory More than 2GB
Display Resolution 1280×1024 pixel
Interface More than 2 USB ports
Application  
Measurement Items Electron Diffusion Coefficients (D) Measurement
Electron Lifetime (τ) Measurement
Electron Density (n) Measurement
I-V Curve Measurement
Data Analysis Creating data table
Drawing Analysis Plot Chart
Short-Circuit Current Density vs. Electron Diffusion Coefficients Measurement
Short-Circuit Current Density vs. Electron Lifetime
Electron Density vs. Electron Lifetime
Electron Density vs. Open-Circuit Voltage